Medfusion
2001 Syringe Pump IV-M-2001
Size: 4.5" X 3" X 7.5", Weight: 2.5 Lbs.
System Accuracy: +/- 3%, Excluding Syringe Variations
Infusion Modes: Continuous, Volume Over Time, Intermittent Auto,
Intermittent Manual Flow Rate:.01 To 378 Ml Per Hour,
Dependent Upon Syringe Size Selected
Syringe Fill Volume: All Syringe Sizes Will Fill To Maximum Stated Volumes
Power: Ac 95-135v, 60hz; Dc Internal Rechargeable Batteries
Recharge Time: 16 Hours At 25º C (With Pump Off)
Battery Capacity: Minimum 10 Hours At 5.0 Ml Per Hour
With A 60 Ml Syringe Total
Volume Delivered: 000.00 To 999.99 Ml Increments Of .01 Ml
Volume Limit: 00.00 To Maximum Capacity Of Syringe Size (With Minimum Programmable
Volume Of 0.05 Ml In Increments Of 0.01 Ml)
Kvo: 0.1 To 9.9 Ml Per Hour In Increments Of 0.1 Ml
(Cannot Exceed Primary Rate)
Medfusion
2010 Syringe Pump
Flow/Rate Range: 0.1-378.0 Ml/Hr Dependent On Syringe Size Selected.
Accuracy: Plus Or Minus 3% (Excluding Syringe Variations,)
Net Weight: 2.5 Lbs.
Power Requirements: Ac Adapter / Dc Internal - Rechargeable Batteries -
Battery Life: 10 Hours At 5.0 Ml/Hr With A 60 Ml Syringe
Alarms: Near Empty, Empty, Volume Limit, Occlusion, System Malfunction,
Low Battery, Depleted Battery, Syringe Not Loaded